Maintenance of spectral, radiometric, and geometric calibration equipment and standards
Betina Pavri
15 December 1999
Order two new NIST-traceable 1000W bulbs (S-style) annually
Order two new 12" and two new 5" spectralon panels annually
Place new standards in locked current calibration cabinet with calibration documentation - only used for AVIRIS calibration
Affix proper mounting hardware to new standards; make sure standard serial numbers are clearly visible
Retire old standards and documentation to secondary cabinet
Enter 5nm splined data for new standards onto networked database
Record hours of use for each bulb on slip in the bulb box
Spectralon reflectance standards must be covered with a clean room (no lint) cloth and covered (to protect them from damage) when not in use
Irradiance standards must be placed back into their boxes when cool
Irradiance standards will be only be handled with gloved hands (clean gloves)
-
| |